Establishment of local diagnostic reference levels for common adult CT examinations: a multicenter survey in Addis Ababa

建立成人常见CT检查的本地诊断参考水平:亚的斯亚贝巴多中心调查

阅读:1

Abstract

BACKGROUND: In medical imaging, a computed tomography (CT) scanner is a major source of ionizing radiation. All medical radiation exposures should be justified and optimized to meet the clinical diagnosis. Thus, to avoid unnecessary radiation doses for patients, diagnostic reference levels (DRLs) have been used. The DRLs are used to identify unusually high radiation doses during CT procedures, which are not appropriate for the clinical diagnosis. It has been successfully implemented in Europe, Canada, Australia, the United States, several industrialized countries, and a few underdeveloped countries. The present study aimed to establish DRLs for the head, chest, and abdominopelvic (AP) CT procedures in Addis Ababa, Ethiopia. METHODS: A pilot study identified the most frequent CT examinations in the city. At the time of the pilot, eighteen CT scan facilities were identified as having functioning CT scanners. Then, on nine CT facilities (50% of functional CT scanners), a prospective analysis of volume CT dose index (CTDI(vol)) and dose length product (DLP) was performed. We collected data for 838 adult patients' head, chest, and AP CT examinations. SPSS version 25 was used to compute the median values of the DLP and CTDI(vol) dose indicators. The rounded 75th percentile of CTDI(vol) and DLP median values were used to define the DRLs. The results are compared to DRL data from the local, regional, and international levels. RESULT: The proposed DRLs using CTDI(vol) (mGy) are 53, 13, and 16 for the head, chest, and AP examinations respectively, while the DLP (mGy.cm) for the respective examinations were 1210, 635, and 822 mGy.cm. CONCLUSION: Baseline CT DRLs figures for the most frequently performed in Addis Ababa were provided. The discrepancies in dose between CT facilities and as well as between identical scanners suggests a large potential for dose optimization of examinations. This can be actually achieved through appropriate training of CT technologists and continuous dose audits.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。