Refined scan protocol for the evaluation of pulmonary perfusion standardized image quality and reduced radiation dose in dynamic chest radiography

改进的扫描方案用于评估肺灌注,可标准化图像质量并降低动态胸部X线摄影中的辐射剂量

阅读:1

Abstract

PURPOSE: Dynamic chest radiography (DCR) is a novel imaging technique used to noninvasively evaluate pulmonary perfusion. However, the standard DCR protocol, which is roughly adapted to the patient's body size, occasionally causes over- or underexposure, which could influence clinical evaluation. Therefore, we proposed a refined protocol by increasing the number of patient body mass index (BMI) categories from three to seven groups and verified its usefulness by comparing the image sensitivity indicators (S-values) and entrance surface doses (ESDs) of the conventional protocol with those of our refined protocol. METHODS: This retrospective observational study included 388 datasets (standing position, 224; supine position, 164) for the conventional protocol (December 2019-April 2021) and 336 datasets (standing position, 233; supine position, 103) for the refined protocol (June-November 2021). The conventional protocol (BMI-3 protocol) divided the patients into three BMI groups (BMI < 17, 17≤BMI < 25, and BMI ≥ 25 kg/m(2) ), whereas the refined protocol (BMI-7 protocol) divided the patients into seven BMI groups (BMI < 17, 17 ≤ BMI < 20, 20 ≤ BMI < 23, 23 ≤ BMI < 26, 26 ≤ BMI < 29, 29 ≤ BMI < 32, and BMI ≥ 32 kg/m(2) ). The coefficients of variation (CVs) for the S-values and ESDs acquired using the two protocols were compared. RESULTS: The CVs of the S-values in the BMI-7 protocol group were significantly lower than those in the BMI-3 protocol group for the standing (28.8% vs. 16.7%; p < 0.01) and supine (24.5% vs. 17.7%; p < 0.01) positions. The ESDs of patients scanned using the BMI-7 protocol were significantly lower than those scanned using the BMI-3 protocol in the standing (1.3 vs. 1.1 mGy; p < 0.01) and supine positions (2.5 vs. 1.6 mGy; p < 0.01), although the mean BMI of the two groups were similar. CONCLUSION: We introduced the BMI-7 protocol and demonstrated its standardized image quality and reduced radiation exposure in patients undergoing DCR.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。