Comprehensive Laboratory Evaluation of a Specific Lateral Flow Assay for the Presumptive Identification of Francisella tularensis in Suspicious White Powders and Aerosol Samples

对可疑白色粉末和气溶胶样本中土拉弗朗西斯菌的推定鉴定的特定横向流动检测法进行全面的实验室评估

阅读:6
作者:Segaran P Pillai, Lindsay DePalma, Kristin W Prentice, Jason G Ramage, Carol Chapman, Jawad Sarwar, Nishanth Parameswaran, Jeannine Petersen, Brook Yockey, John Young, Ajay Singh, Christine A Pillai, Gowri Manickam, Nagarajan Thirunavkkarasu, Julie R Avila, Shashi Sharma, Stephen A Morse, Kodumudi V

Abstract

We conducted a comprehensive, multi-phase laboratory evaluation of the Tularemia BioThreat Alert® (BTA) test, a lateral flow assay (LFA) for the rapid detection of Francisella tularensis. The study, conducted at 2 sites, evaluated the limit of detection (LOD) of this assay using the virulent SchuS4 strain and the avirulent LVS strain of F. tularensis. In 6-phase evaluation (linear dynamic range and reproducibility, inclusivity, near-neighbor, environmental background, white powder, and environmental filter extract), 13 diverse strains of F. tularensis, 8 Francisella near neighbors, 61 environmental background organisms, 26 white powders, and a pooled aerosol extract were tested. In the 937 tests performed, the Tularemia BTA demonstrated an LOD of 107 to 108 cfu/mL, with a sensitivity of 100.00%, specificity of 98.08%, and accuracy of 98.84%. These performance data are important for accurate interpretation of qualitative results arising from screening suspicious white powders in the field.