Evaluation of the Copan ESwab Transport System for Viability of Pathogenic Fungi by Use of a Modification of Clinical and Laboratory Standards Institute Document M40-A2

采用临床和实验室标准协会文件 M40-A2 的改进方法评估 Copan ESwab 运输系统对致病真菌的存活率

阅读:1

Abstract

The Copan ESwab system was evaluated for its ability to maintain the viability of pathogenic fungi. Tests followed the Clinical and Laboratory Standards Institute (CLSI) document M40-A2 roll-plate method at room and refrigerator temperatures. A system was devised for standardizing homogeneous inoculum suspensions of variously sized conidia and sporangiospores of filamentous fungi. A total of 19 clinical and reference strains were standardized to a 0.5 McFarland turbidity standard with a simple photometer. Corresponding optical densities were measured with a spectrophotometer. Colony counts equal to or greater than those seen at time zero were obtained for the entire test panel. Results indicate that the Copan ESwab system effectively maintains prevalent opportunistic fungal organisms for at least 48 h.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。