A Novel Approach to Evaluating Crosstalk for Near-Infrared Spectrometers

一种评估近红外光谱仪串扰的新方法

阅读:1

Abstract

Multi-channel and multi-parameter near-infrared spectroscopy (NIRS) has gradually become a new research direction and hot spot due to its ability to provide real-time, continuous, comprehensive indicators of multiple parameters. However, multi-channel and multi-parameter detection may lead to crosstalk between signals. There is still a lack of benchmarks for the evaluation of the reliability, sensitivity, stability and response consistency of the NIRS instruments. In this study, a set of test methods (a human blood model test, ink drop test, multi-channel crosstalk test and multi-parameter crosstalk test) for analyzing crosstalk and verifying the reliability of NIRS was conducted to test experimental verification on a multi-channel (8-channel), multi-parameter (4-parameter) NIRS instrument independently developed by our team. Results show that these tests can be used to analyze the signal crosstalk and verify the reliability, sensitivity, stability and response consistency of the NIRS instrument. This study contributes to the establishment of benchmarks for the NIRS instrument crosstalk and reliability testing. These novel tests have the potential to become the benchmark for NIRS instrument reliability testing.