Multimodal Neuroelectrophysiological Monitoring Combined with Robot-Assisted Placement of a Transiliac-Transsacral Screw for the Treatment of Transforaminal Sacral Fractures

多模式神经电生理监测联合机器人辅助经髂骶螺钉置入术治疗经椎间孔骶骨骨折

阅读:2

Abstract

OBJECTIVE: This study aimed to evaluate the safety and efficacy of the fixation of transforaminal sacral fractures using TiRobot-assisted transiliac-transsacral (TITS) screws under multimodal neuroelectrophysiological monitoring (MNM). METHODS: From January 2019 to May 2021, 22 patients (17 male and 5 female patients) with transforaminal sacral fractures who were treated with closed reduction and placement of TiRobot-assisted TITS screws under MNM were retrospectively evaluated. The average age of the patients was 43.32 ± 11.40 years (range: 19-63). The patients received MNM, including somatosensory-evoked potentials (SEPs), motor-evoked potentials (MEPs), and electromyographic monitoring (EMG), prior to surgery, during closed reduction and the placement of the guidewire and TITS screw, and at the end of surgery. The operation was adjusted according to the MNM results. RESULTS: Overall, 22 TITS screws were inserted in 22 patients, including 5 TITS screws in the S1 body and 17 TITS screws in the S2 body. The average time needed for screw placement was 27.95 ± 6.84 mins, and the average frequency of X-ray fluoroscopy exposures was 31.00 ± 5.56 for each patient. Anterior ring fixation was performed in 4 patients using an external fixator, in 5 patients using cannulated screws, and in 13 patients using reconstruction plates. The mean follow-up time was 14.46 ± 2.46 months (12-20 months). Tornetta and Matta radiographic outcomes were excellent in 10 patients, good in 9 patients, fair in 2 patients, and poor in 1 patient. The proportion of excellent and good ratings was 86.36%. At the final follow-up, the average Majeed score was 82.18 ± 14.52, with clinical outcomes that were excellent in 9 patients, good in 9 patients, fair in 1 patient, and poor in 3 patients. The proportion of excellent and good ratings was 82.82%. Preoperatively, the amplitude of the SEP on the injured side was lower than that on the contralateral side before reduction in 9 patients (>50%). In this study, no screw was mistakenly inserted into the sacral canal, and no surgical site infection occurred. CONCLUSION: MNM combined with TiRobot assistance can safely implant TITS screws and can effectively identify the neurological function of patients under anesthesia and reduce iatrogenic nerve injury.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。