Computational FEM Model and Phantom Validation of Microwave Ablation for Segmental Microcalcifications in Breasts Using a Coaxial Double-Slot Antenna

利用同轴双缝天线进行微波消融治疗乳腺节段性微钙化的计算有限元模型和体模验证

阅读:1

Abstract

INTRODUCTION: Cancer is the second leading cause of death worldwide. Breast cancer is the second most common cause of cancer-related mortality, accounting for 11.6% of the total number of deaths. The main treatments for this disease are surgical removal of the tumor, radiotherapy, and chemotherapy. Recently, different minimally invasive technologies have been applied (e.g., emission of electromagnetic waves, thermal and chemical means) to overcome the important side effects of these treatment modalities. The objective of this study was to develop and evaluate a predictive computational model of microwave ablation. MATERIALS AND METHODS: The predictive computational model of microwave ablation was constructed by means of a dual-slot coaxial antenna. The model was compared with an experiment performed using a breast phantom, which emulates the dielectric properties of breast tissue with segmental microcalcifications. The standing wave ratio (SWR) was obtained for both methods to make a comparison and determine the feasibility of applying electromagnetic ablation to premalignant lesions in breasts. Specifically, for the analysis of segmental microcalcifications, a breast phantom with segmental microcalcifications was developed and two computational models were performed under the same conditions (except for blood perfusion, which was excluded in one of the models). RESULTS: The SWR was obtained by triplicate experiments in the phantom, and the measurements had a difference of 0.191 between the minimum and maximum SWR values, implying a change of power reflection of 0.8%. The average of the three measurements was compared with the simulation that did not consider blood perfusion. The comparison yielded a change of 0.104, representing a 0.2% change in power reflection. Discussion. Both experimentation in phantom and simulations demonstrated that ablation therapy can be performed using this antenna. However, an additional optimization procedure is warranted to increase the efficiency of the antenna.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。