3D Printed Guides and Preoperative Planning for Uncemented Stem Anteversion Reconstruction during Hip Arthroplasty: A Pilot Study

3D打印导板和术前规划在髋关节置换术中用于无骨水泥柄前倾角重建:一项初步研究

阅读:2

Abstract

OBJECTIVE: To investigate if 3D printed guides and preoperative planning can accurately control femoral stem anteversion. METHODS: A prospective comparative study was carried out from 2018 to 2020, including 53 patients who underwent hip arthroplasty for femoral neck fracture. The target rotation center of the femoral head is determined by three-dimensional planning. In group A, planning was made by 2D templates. In group B, preoperative 3D planning and 3D printed osteotomy/positioning guides were performed. After the operation, 3D model registration was performed to calculate the accuracy of anteversion restoration. RESULTS: We screened 60 patients and randomized a total of 53 to 2 parallel study arms: 30 patients to the group A (traditional operation) and 23 patients to the group B (3D preoperative planning and 3D printed guide). There were no significant differences in demographic or perioperative data between study groups. The restoration accuracy of group A was 5.42° ± 3.65° and of group B was 2.32° ± 1.89°. The number and rate of abnormal cases was 15 (50%) and 2 (8.7%), respectively. Significant statistical differences were found in angle change, restoration accuracy, and number of abnormal cases. CONCLUSION: Three-dimensional preoperative planning and 3D printed guides can improve the accuracy of the restoration of femoral anteversion during hip arthroplasty.