A New Silicon Dioxide-Coated MALDI-ToF Sample Plate for Peptide Analysis

一种用于肽分析的新型二氧化硅涂层MALDI-ToF样品板

阅读:1

Abstract

In this report, we describe the development and testing of a new coated plate which improves the sensitivity and accuracy in matrix-assisted laser desorption/ionization time-of-flight mass spectrometry (MALDI-ToF-MS). The coated plate was covered with a thin layer of hydrophobic silicon dioxide, which enabled sample enrichment due to the water repellent nature of the silicon dioxide surface. Sensitivity and required laser strengths were tested using peptide standards, with the results that these coated plates required lower laser power and showed increased sensitivity than that of common plates. Accuracy was tested using bacteria, saliva, and serum samples. The coated plates showed significantly increased degrees of accuracy through their capacity to reduce mass shift. The importance and necessity of accuracy analysis in the assessment of new sample plates, which is rarely described in other papers, is also discussed.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。