Influence of Irradiation Temperature on Oxidative and Network Properties of X-Ray Cross-Linked Vitamin E Stabilized UHMWPE for Hip Arthroplasty

辐照温度对用于髋关节置换术的X射线交联维生素E稳定超高分子量聚乙烯的氧化和网络特性的影响

阅读:1

Abstract

Previous studies have shown that increased cross-link density, reduced free radicals, and increased antioxidant grafting resulting from electron-beam irradiation at elevated temperatures improved the wear performance and the oxidative stability of vitamin E blended UHMWPE. The current study explores the impact of elevated irradiation temperature on vitamin E blended UHMWPE using X-ray. We hypothesize that the effects of temperature would be similar to those observed after electron-beam irradiation due to the relatively high dose rate of X-rays. Two X-ray doses of 80 and 100 kGy and two irradiation temperatures, that is, room temperature and 100°C were considered. The reference was Vitelene®, a vitamin E stabilized polyethylene cross-linked with 80 kGy by e-beam at 100°C. Oxidation index and oxidation induction time, as well as cross-link density, gel fraction, and trans-vinylene index, were determined, as the oxidative and network properties are decisive for the long-term implant performance. Gel fraction and oxidation induction time were significantly improved subsequently to warm irradiation in comparison with the material irradiated at room temperature. In conclusion, X-ray irradiation at elevated temperatures resulted in an increase of cross-linking and oxidative resistance of vitamin E stabilized polyethylene comparable to those of e-beam irradiated UHMWPE.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。