Comparative Evaluations on Real-Time Monitoring of Temperature Sensors during Endoscopic Laser Application

内窥镜激光应用中温度传感器实时监测的比较评价

阅读:1

Abstract

Temperature sensors, such as Fiber Bragg Grating (FBG) and thermocouple (TC), have been widely used for monitoring the interstitial tissue temperature during laser irradiation. The aim of the current study was to compare the performance of both FBG and TC in real-time temperature monitoring during endoscopic and circumferential laser treatment on tubular tissue structure. A 600-µm core-diameter diffusing applicator was employed to deliver 980-nm laser light (30 W for 90 s) circumferentially for quantitative evaluation. The tip of the TC was covered with a white tube (W-TC) in order to prevent direct light absorption and to minimize temperature overestimation. The temperature measurements in air demonstrated that the measurement difference in the temperature elevations was around 3.5 °C between FBG and W-TC. Ex vivo porcine liver tests confirmed that the measurement difference became lower (less than 1 °C). Ex vivo porcine esophageal tissue using a balloon-integrated catheter exhibited that both FBG and W-TC consistently showed a comparable trend of temperature measurements during laser irradiation (~2 °C). The current study demonstrated that the white tube-covered TC could be a feasible sensor to monitor interstitial tissue temperature with minimal overestimation during endoscopic laser irradiation. Further in vivo studies on gastroesophageal reflux disease will investigate the performance of the W-TC to monitor the temperature of the esophageal mucosa surface in real-time mode to warrant the safety of endoscopic laser treatment.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。