Verification of patient-setup accuracy using a surface imaging system with steep measurement angle

利用具有陡峭测量角度的表面成像系统验证患者摆位精度

阅读:2

Abstract

PURPOSE: We evaluate an SGRT device (Voxelan HEV-600 M/RMS) installed with Radixact, with the view angle of the Voxelan's camera at 74 degrees. The accuracy of Voxelan with this steep angle was evaluated with phantom experiments and inter-fractional setup errors of patients. METHODS: In the phantom experiments, the difference between the measured values of Voxelan from the truth was evaluated for translations and rotations. The inter-fractional setup error between the setup using skin markers with laser localizer (laser setup: LS) and the setup using Voxelan (surface setup: SS) was compared for head and neck (N = 19), chest (N = 7) and pelvis (N = 9) cases. The inter-fractional setup error was calculated by subtracting from bone matching by megavoltage computed tomography (MVCT) as ground truth. RESULTS: From the phantom experiments, the average difference between the measured values of Voxelan from the truth was within 1 mm and 1 degree. In all cases, inter-fractional setup error based on MVCT was not significantly different between LS and SS by Welch's t-test (P > 0.05). The vector offset of the LS for head and neck, chest, and pelvis were 6.5, 9.6, and 9.6 mm, respectively, and that of the SS were 5.8, 8.6, and 12.6 mm, respectively. Slight improvement was observed for the head and neck, and chest cases, however, pelvis cases were not improved because the umbilical region could not be clearly visualized as a reference. CONCLUSION: The results show that SS in Voxelan with an installation angle of 74 degrees is equal to or better than LS.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。