A novel blood pressure monitoring technique by smart HUAWEI WATCH: A validation study according to the ANSI/AAMI/ISO 81060-2:2018 guidelines

基于华为手表的新型血压监测技术:根据 ANSI/AAMI/ISO 81060-2:2018 指南的验证研究

阅读:1

Abstract

BACKGROUND: Given the rapid innovation of wearable technology, additional physical indicators can be detected, and blood pressure (BP) has become the focus of many emerging medical-device manufacturers. This study aimed to validate the accuracy of the newly developed HUAWEI WATCH in BP monitoring, according to the American National Standards Institute/Association for the Advancement of Medical Instrumentation/International Organization for Standardization (ANSI/AAMI/ISO 81060-2:2018) guidelines. MATERIALS AND METHODS: The same arm sequential BP measurement was applied. One validation included four reference BP measurements taken simultaneously by two independent observers using a mercury sphygmomanometer, alternating with three test-watch measurements. Each test-watch measurement was compared against the average of the previous and subsequent reference BP readings. Two criteria were required for validation: (1) a mean BP difference of 5 mm Hg or less, with a standard deviation (SD) of 8 mm Hg or less for systolic blood pressure (SBP) and diastolic blood pressure (DBP) in the 255 pairs of measurements, and (2) an SD for the of 85 averaged BP differences within the threshold defined by the mean test-reference BP difference listed in the ANSI/AAMI/ISO 81060-2:2018 guidelines. RESULTS: The mean age of the 85 participants was 48 ± 18 years (range: 21-85), and 53 (62.4%) were male. The mean differences between the test and reference BPs were -0.25 ± 5.62 mm Hg and -1.33 ± 6.81 mm Hg for SBP and DBP, respectively (according to Criterion 1). The mean differences between the test BPs and reference BPs were -0.25 ± 5.00 mm Hg and -1.33 ± 6.31 mm Hg for SBP and DBP, respectively, according to Criterion 2. CONCLUSION: Blood pressure measurement using the HUAWEI WATCH showed excellent consistency with reference BPs, and fulfilled both validation criteria of the guidelines, show its promise as a wearable device for BP self-monitoring.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。