Radiation exposure and screening yield by digital breast tomosynthesis compared to mammography: results of the TOSYMA Trial breast density related

数字乳腺断层合成与乳房 X 线摄影的辐射暴露和筛查效果比较:TOSYMA 试验结果与乳腺密度相关

阅读:12
作者:Alexander Sommer, Stefanie Weigel, Hans-Werner Hense, Joachim Gerß, Veronika Weyer-Elberich, Laura Kerschke, Elke Nekolla, Horst Lenzen, Walter Heindel; TOSYMA Screening Trial Study Group

Conclusion

Women with dense breasts may specifically benefit from DBT + SM screening as high cancer detection is achieved with only moderate AGD elevations. Clinical relevance statement: TOSYMA suggests a favorable constellation for screening with digital breast tomosynthesis plus synthesized mammography (DBT + SM) in dense breasts when weighing average glandular dose elevation against raised invasive breast cancer detection rates. There is potential for density-, i.e., risk-adapted population-wide breast cancer screening with DBT + SM. Key points: Breast thickness declines with visually increasing density in digital mammography (DM) and digital breast tomosynthesis (DBT). Average glandular doses of DBT decrease with increasing density; digital mammography shows lower and more constant values. With the smallest average glandular dose difference in dense breasts, DBT plus SM had the highest difference in invasive breast cancer detection rates.

Methods

TOSYMA screened 99,689 women aged 50 to 69 years. Compression force, resulting breast thickness, the calculated AGD obtained from each mammography device, and previously published iCDR were used for comparisons across breast density categories in the two trial arms.

Results

There were 196,622 exposures of 49,227 women (DBT + SM) and 197,037 exposures of 49,132 women (DM) available for analyses. Mean breast thicknesses declined from breast density category A (fatty) to D (extremely dense) in both trial arms. However, while the mean AGD in the DBT + SM arm declined concomitantly from category A (2.41 mGy) to D (1.89 mGy), it remained almost unchanged in the DM arm (1.46 and 1.51 mGy, respectively). In relative terms, the AGD elevation in the DBT + SM arm (64.4% (A), by 44.5% (B), 27.8% (C), and 26.0% (D)) was lowest in dense breasts where, however, the highest iCDR were observed.