Evaluation of the accuracy of two point-of-care haemoglobin meters used in sub-Saharan African population: a cross-sectional study

评估两种用于撒哈拉以南非洲人群的床旁血红蛋白仪的准确性:一项横断面研究

阅读:1

Abstract

BACKGROUND: Point-of-care haemoglobin meters are attractive solutions to improve timely diagnosis of anaemia in resource-limited settings. However, concerns regarding the accuracy of these meters may affect their adoption. The accuracy of two hand-held point-of-care haemoglobin meters was evaluated against reference full blood count analyser. METHODS: This was a hospital-based cross-sectional study conducted at the Douala General hospital, Cameroon. Two handheld haemoglobin meters were assessed: Urit12® (URIT Medical Electronics Co.,Ltd. Guangxi, China) and MissionHb®(ACON Laboratories, Inc., San Diego, USA); against a reference standard CELL-DYN RUBY® (ABBOTT DIAGNOSTICS, Illinois, USA). The Pearson's correlation and Bland-Altman agreement were used to assess the technical accuracy of the meters. Clinical accuracy was evaluated using total error allowable and area under the Receiver Operating Curve. Finally, their agreement with the reference in diagnosing anaemia was assessed using the kappa statistic. RESULTS: A total of 228 participants were included in the study. The mean haemoglobin values of both haemoglobin meters (MissionHb®: 11.6 ± 2.5 g/dl; Urit12®: 10.9 ± 2.7 g/dl) were significantly higher than the reference value (10.5 ± 2.5 g/dl), p < 0.001 for both meters. Both haemoglobin meters had good correlation with the reference analyser (r = 0.89 and r = 0.90 for Urit12® and MissionHb® respectively) and good agreement on the Bland-Altman plots. However, the MissionHb® meter did not meet the clinical accuracy requirements (p < 0.001). Even though both meters were excellent at identifying the presence of anemia (MissionHb® - AUC = 0.9161, Urit 12® - AUC = 0.9009), they, however, both had weak agreement with the reference analyser in diagnosing the severity of anaemia (K = 0.39 for MissionHb®, p < 0.001 and K = 0.54 for Urit12®, p < 0.001). CONCLUSION: Although both devices showed technical accuracy with a positive correlation with the reference analyser and were able to accurately diagnose the presence of anemia, both meters however, had sub-optimal agreement with the reference analyser in diagnosing the degree of severity of anaemia among our participants.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。