Factors influencing the accuracy of non-invasive blood pressure measurements in patients admitted for cardiogenic shock

影响心源性休克患者无创血压测量准确性的因素

阅读:2

Abstract

BACKGROUND: Although invasively measured blood pressure (invBP) is regarded as a "gold standard" in critically ill cardiac patients, the non-invasive BP is still widely used, at least at the initiation of medical care. The erroneous interpretation of BP can lead to clinical errors. We therefore investigated the agreement of both methods with respect to some common clinical situation. METHODS: We included 85 patients hospitalized for cardiogenic shock. We measured BP every 6 h for the first 72 h of hospitalization, in all patients. Each set of BP measurements included two invasive (invBP), two auscultatory (auscBP), and two oscillometric (oscBP) BP measurements. InvBP was considered as a gold standard. Mean non-invasive arterial pressure (MAP) was calculated as (diastolic pressure + (pulse pressure ÷ 3)). We used Bland-Altman analysis and we calculated concordance correlation coefficients to assess agreement between different BP methods. RESULTS: We obtained 967 sets of BP measurements. AuscMAP and oscMAP were on average only 0.4 ± 8.2 and 1.8 ± 8.5 mmHg higher than invMAP, respectively. On the other hand, auscSBP and oscSBP were on average - 6.1 ± 11.4 and - 4.1 ± 9.8 mmHg lower than invSBP, respectively. However, the mean differences and variability for systolic and diastolic BP variability were large; the 2 standard deviation differences were ± 24 and 18 mmHg. In hypotension, non-invasive BP tended to be higher than invBP while the opposite was true for high BP values. Clinical conditions associated with hypotension generally worsened the accuracy of non-invasive MAP. CONCLUSIONS: Mean arterial pressure measured non-invasively appears to be in good agreement with invasive MAP in patients admitted for cardiogenic shock. Several clinical associated with hypotension can affect accuracy of non-invasive measurement. Auscultatory and oscillometric measurements had similar accuracy even in patients with arrhythmia.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。