An inappropriate pacing threshold increase after repeated electrical storm in a patient with implantable cardioverter defibrillator

植入式心脏复律除颤器患者在反复发生电风暴后出现不适当的起搏阈值升高

阅读:2

Abstract

BACKGROUND: Implantable cardioverter defibrillators (ICD) are capable of effectively terminating malignant ventricular arrhythmia and are the most effective way to prevent sudden cardiac death. However, some evidences demonstrated that both anti-tachycardia pacing (ATP) and ICD shock can also bring adverse prognosis. CASE PRESENTATION: A 66-year-old Han Chinese man with prior ICD implantation was admitted to our hospital because of frequent ICD shocks. Although intravenous amiodarone and esmolol succinate were administered daily, the patient suffered 155 episodes of VT/VF during 8 weeks after implantation. After repeated discharge of the device, the pacing threshold of the patient increased gradually. Considering the inappropriate increase of the pacing threshold, we decided to reposition the right ventricular (RV) lead with good sensing and threshold parameters confirmed. Subsequent 22 months interrogation follow-up revealed a stable lead position and electrical specifications. Furthermore, antiarrhythmic drugs were maximally increased, while ATP burst was remarkably decreased and the inappropriate ICD shock never occurred until now. CONCLUSION: An inappropriate pacing threshold was increased secondary to repeated ICD electrical storm. A timely active lead position adjustment reduced the pacing threshold and eliminated the risk of premature battery depletion.