Efficacy of a seal-wing paclitaxel-eluting balloon catheters in the treatment of bare metal stent restenosis

密封翼紫杉醇洗脱球囊导管治疗裸金属支架再狭窄的疗效

阅读:2

Abstract

BACKGROUND: Our study aimed to compare the efficacy of seal-wing paclitaxel-eluting balloon catheters (PEB) with iopromide-coated PEB and everolimus-eluting stents (EES) for treating bare metal stent restenosis (BMS-ISR). METHODS: We enrolled 64 patients with 69 BMS-ISR. The control group comprised patients from the iopromide-PEB and EES arms of a previous TIS study. The primary end-point was 12-month in-segment late lumen loss (LLL). Secondary end-points included incidence of binary in-stent restenosis and 12-month major adverse cardiac events (MACE). RESULTS: Compared to iopromide-coated PEB, seal-wing PEB was associated with significantly higher 12-month LLL (0.30 vs. 0.02 mm; p < 0.0001), repeated binary restenosis (28.12% vs. 8.7%; p = 0.012), 12-month MACE (26.98% vs. 10.29%; p = 0.003), and target vessel revascularization (TVR; 20.63% vs. 7.35%; p = 0.009). Compared to EES, no significant differences were found in the 12-month LLL (0.30 vs. 0.19 mm; p = 1.000), repeated binary restenosis (28.12% vs. 19.12%; p = 0.666), 12-month MACE (26.98% vs. 19.12%; p = 0.102) or TVR (20.63% vs. 16.18%; p = 0.360). CONCLUSION: BMS-ISR treatment using seal-wing PEB led to significantly higher 12-month LLL, repeated binary restenosis, MACE, and TVR compared to iopromide-coated PEB. However, no significant differences were found in comparison with EES. TRIAL REGISTRATION: ClinicalTrials.gov; NCT01735825.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。