Quantitative evaluation of microvacuole formation in five intraocular lens models made of different hydrophobic materials

对五种由不同疏水材料制成的人工晶状体模型中微空泡的形成进行定量评估

阅读:2

Abstract

In this laboratory study, we assessed the resistance to microvacuole (glistening) formation in hydrophobic intraocular lenses (IOLs). Glistenings were induced in five lenses each of five different hydrophobic acrylic IOL models, using an established in vitro laboratory model: 800C (Rayner, Worthing, UK), AcrySof SN60WF (Alcon, Fort Worth, USA), Tecnis ZCB00 (Johnson & Johnson Vision, Santa Ana, USA), Vivinex XY1 (Hoya, Tokyo, Japan) and CT Lucia 611P (Zeiss, Oberkochen, Germany). We evaluated the number of microvacuoles per square millimeter (MV/mm2) in the central part of each IOL. Results were analyzed statistically, and mean glistening numbers were ranked, with the highest in the SN60WF which had 66.0 (±45.5) MVs/mm, followed by the 611P with 30.7 (±8.4) MVs/mm2. The 800C and XY1 showed comparable values of 2.0 (±3.6) and 2.7 (±2.4) MVs/mm2, respectively. ZCB00 had the lowest number with 0.9 (±0.6) MVs/mm2. This study shows that the resistance to glistening formation differs depending on the hydrophobic acrylic copolymer composition of the IOL material. Some IOLs from current clinical use are still prone to develop glistenings whereas others, including the ZCB00, 800C and XY1 show high resistance to microvacuole formation.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。