MIRD Pamphlet No. 32: A MIRD Recovery Coefficient Model for Resolution Characterization and Shape-Specific Partial-Volume Correction

MIRD 手册第 32 号:用于分辨率表征和形状特定部分体积校正的 MIRD 恢复系数模型

阅读:1

Abstract

Accurate quantification in emission tomography is essential for internal radiopharmaceutical therapy dosimetry. Mean activity concentration measurements in objects with diameters less than 10 times the full width at half maximum of the imaging system's spatial resolution are significantly affected (>10%) by the partial-volume effect. This study develops a framework for PET and SPECT spatial resolution characterization and proposes 2 MIRD recovery coefficient models-a geometric mean approximation (RECOVER-GM) and an empirical model (RECOVER-EM)-that provide shape-specific partial-volume correction (PVC). The models were validated using simulations and phantom experiments, with a comparative PVC test on ellipsoidal phantoms demonstrating that the RECOVER models significantly reduced error in activity quantification by factors of approximately 1.3-5.7 compared with conventional sphere-based corrections. The proposed recovery coefficient models and PVC methodology provide a robust framework for improved region-based PVC, including corrections for nonspherical tumor volumes. This work is part of the ongoing MIRDsoft.org project that aims to enhance accessibility to advanced dosimetry tools for improved disease characterization, treatment planning, and radiopharmaceutical therapy dosimetry.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。