Considering the Node Level in Error Correction for DMFBs

考虑DMFB纠错中的节点级

阅读:1

Abstract

In recent years, a type of biochip known as a Digital Microfluidic Biochip (DMFB) has been actively researched in the field of life sciences. DMFBs perform dilution operations by mixing reagent solutions and buffer solutions at a 1:1 ratio to generate droplets with the desired concentration. One of the challenges of DMFBs is that droplets may not always be evenly split during the droplet division process. To address this issue, an error correction method utilizing error cancellation has been proposed. This method modifies the dilution graph to minimize the impact of division errors on the target node. However, this approach has a significant drawback: when large division errors occur in nodes close to the target node, they can introduce substantial concentration errors at the target node. In this paper, we propose a method that duplicates nodes near the target node and performs re-dilution to correct errors. Furthermore, we present an efficient and accurate error correction approach by modifying the dilution graph so that the output nodes of the dilution operation are at equal levels relative to the target node. Through simulations conducted 10,000 times, we demonstrate that our method effectively reduces the average concentration error at the target node.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。