Dynamic Error Correction of Filament Thermocouples with Different Structures of Junction based on Inverse Filtering Method

基于逆滤波法的不同结结构丝状热电偶动态误差校正

阅读:1

Abstract

Since filament thermocouple is limited by its junction structure and dynamic characteristics, the actual heat conduction process cannot be reproduced during the transient thermal shock. In order to solve this problem, we established a thermocouple dynamic calibration system with laser pulse as excitation source to transform the problem of the restoring excitation source acting on the surface temperature of thermocouple junction into the problem of solving the one-dimensional (1D) inverse heat conduction process, proposed a two-layer domain filtering kernel regularization method for double conductors of thermocouple, analyzed the factors causing unstable two-layer domain solution, and solved the regular solution of two-layer domain by the filtering kernel regularization strategy. By laser narrow pulse calibration experiment, we obtained experimental samples of filament thermocouples with two kinds of junction structures, butt-welded and ball-welded; established error estimation criterion; and obtained the optimal filtering kernel parameters by the proposed regularization strategy, respectively. The regular solutions solved for different thermocouples were very close to the exact solution under the optimal strategy, indicating that the proposed regularization method can effectively approach the actual surface temperature of the thermocouple junction.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。