Multicentre evaluation of the BYG Carba v2.0 test, a simplified electrochemical assay for the rapid laboratory detection of carbapenemase-producing Enterobacteriaceae

BYG Carba v2.0 检测的多中心评估,该检测是一种简化的电化学检测方法,用于快速实验室检测产碳青霉烯酶肠杆菌科细菌。

阅读:1

Abstract

Rapid detection of carbapenemase-producing Enterobacteriaceae (CPE) represents a major challenge for microbiology laboratories. We evaluated the BYG Carba v2.0 using a simplified protocol, which detects CPE in less than 30 minutes. This new procedure reduces the hands-on-time from 5 to one minute and only requires a limited amount of material (one to three colonies) thereby preventing the need for subculturing bacterial isolates to reach a larger amount of pure biomass. This multicentre study involved four European reference laboratories. For the 1181 isolates tested across four centres, BYG Carba v2.0 yielded overall sensitivity and specificity of 96.3% (CI95: 94.5-97.5) and 99.7% (CI95: 98.6-100) respectively. Considering only the 670 consecutive isolates tested prospectively, the BYG Carba v2.0 displayed overall positive and negative predictive values of 99.7% (CI95: 95.4-98.9) and 97.5% (CI95: 94.9-98.8). Regarding time to positivity, 85% of CPE detected were positive within ten minutes. The BYG Carba v2.0 is a new highly simplified, rapid and accurate electrochemical assay discriminating between CPE and non-CPE in less than 30 min. The real-time quantified signal allows objective and traceable interpretation of the results.