A New Calibration Circuit Design to Reduce Drift Effect of RuO(2) Urea Biosensors

一种降低RuO(2)尿素生物传感器漂移效应的新型校准电路设计

阅读:1

Abstract

The goal of this study was to reduce the drift effect of RuO(2) urea biosensors. A new calibration circuit (NCC) based on the voltage regulation technique with the advantage of having a simple structure was presented. To keep its simplicity, the proposed NCC was composed of a non-inverting amplifier and a voltage calibrating circuit. A ruthenium oxide (RuO(2)) urea biosensor was fabricated to test the calibrating characteristics of the drift rate of the proposed NCC. The experiment performed in this study was divided into two main stages. For the first stage, a sound RuO(2) urea biosensor testing environment was set-up. The RuO(2) urea sensing film was immersed in the urea solution for 12 h and the response voltage was measured using the voltage-time (V-T) measurement system and the proposed NCC. The results of the first stage showed that the RuO(2) urea biosensor has an average sensitivity of 1.860 mV/(mg/dL) and has a linearity of 0.999 which means that the RuO(2) urea biosensor had been well fabricated. The second stage of the experiment verified the proposed NCC's functions, and the results indicated that the proposed NCC reduced the drift rate of RuO(2) urea biosensor to 0.02 mV/hr (98.77% reduction).