Development of electrical substitution Fourier transform spectrometry for absolute optical power measurements

开发用于绝对光功率测量的电替代傅里叶变换光谱法

阅读:1

Abstract

We have demonstrated the first continuous-scan electrical substitution Fourier transform spectrometer (ES-FTS), which serves initially as an apparatus for absolute spectral responsivity calibrations of detectors over the wavelength range from 1.5 µm to 11 µm. We present data on the realization of a spectral detector-comparator system with high accuracy, high dynamic range, high spectral resolution and fast measurement in the infrared region, which is tied directly to an absolute power scale through electrical substitution. The ES-FTS apparatus employs a commercial Fourier transform spectrometer and a custom electrical substitution bolometer detector to enable spectrally-resolved absolute optical power measurements. A generalization of electrical substitution techniques enables determination of the voltage waveform that must be applied to the bolometer's electrical heater to cancel the optical signal from a Michelson interferometer in order to quantify the time-dependent optical power incident on the bolometer. The noise floor of the electrical substitution bolometer is on the order of 10 pW/Hz(½) and its response is expected to be linear from the noise floor to 1 mW. A direct comparison between a pyroelectric standard detector and the ES-FTS has been performed, and experimental results reported here show great potential for this technique.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。