Compressed sensing-enabled phase-sensitive swept-source optical coherence tomography

基于压缩感知的相位敏感扫频源光学相干断层扫描

阅读:1

Abstract

Here we present a novel phase-sensitive swept-source optical coherence tomography (PhS-SS-OCT) system. The simultaneously recorded calibration signal, which is commonly used in SS-OCT to stabilize the phase, is randomly sub-sampled during the acquisition, and it is later reconstructed based on the Compressed Sensing (CS) theory. We first mathematically investigated the method, and verified it through computer simulations. We then conducted a vibrational frequency test and a flow velocity measurement in phantoms to demonstrate the system's capability of handling phase-sensitive tasks. The proposed scheme shows excellent phase stability with greatly discounted data bandwidth compared with conventional procedures. We further showcased the usefulness of the system in biological samples by detecting the blood flow in ex vivo swine left marginal artery. The proposed system is compatible with most of the existing SS-OCT systems and could be a preferred solution for future high-speed phase-sensitive applications.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。