Multicenter evaluating of a commercially available PCR assay for diagnosing enterovirus infection in a panel of cerebrospinal fluid specimens

对市售PCR检测方法在脑脊液样本中诊断肠道病毒感染的多中心评估

阅读:1

Abstract

Thirteen laboratories participated in blind tests of a panel of 20 coded cerebrospinal fluid specimens (7 uninfected samples, 3 samples infected with 1 50% tissue culture infective dose [TCID50]/0.1 ml [nonenterovirus strains], and 10 samples infected with 10, 1, or 0.1 TCID50/0.1 ml [three different enterovirus serotypes]) on the Amplicor enterovirus PCR assay (Roche Diagnostic Systems). The panel was also evaluated by in-house PCR (two nested-PCR and three one-step PCR assay) or tissue culture (eight laboratories). The viral load was shown to influence greatly the sensitivity of the assay. The average sensitivity of the Amplicor test ranged from 67 to 98% for viral titers of 1 to 10 TCID50/0.1 ml, respectively; titers of 0.1 TCID50/0.1 ml resulted in a sensitivity of only 16%. The overall specificity of the Amplicor test was 98%. The Amplicor assay compared favorably to the five in-house PCR tests (no significant difference in either sensitivity or specificity) and was much more sensitive than tissue culture (P < 0.001), even for high viral loads. It was easy to perform, rapid (about 6 h), well-standardized, and appeared to be suitable for the diagnosis of enterovirus meningitis on a routine basis in laboratories trained in molecular biology techniques.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。