Effects of changes in the waveform and frequency of radio frequency energy on tissue ablation range

射频能量波形和频率变化对组织消融范围的影响

阅读:1

Abstract

This study reports the effects of changes in the waveform and frequency of radio frequency (RF) energy on the tissue ablation range. We developed a 70-watt RFA generator that provides sine and square waves and allows frequency control between 10 Hz and 500 kHz. The changes in the ablation range according to the waveform and frequency were observed using the developed generator. In the waveform variation test, the distance between the electrodes and the electrode type were changed for both waveforms with the frequency set to 500 kHz. In the frequency variation test, the waveform and electrode type were changed with the frequency set to 10, 100, and 500 kHz, while the distance between the electrodes was set to 20 mm. A fixed 45 voltage was applied using the bipolar method. RF energy was applied for 90 s in vitro. The temperature was regulated to not exceed 70°C. The ablation range was calculated using ImageJ software. The analysis results showed that the ablation range was larger with the square wave than with the sine wave and at 10 kHz than at 500 kHz. The developed generator can advance research on ablation area and depth in RF ablation.