Parameter Calibration of Rotating Wave Plate Polarization Detection Device Using Dual Beams

利用双光束对旋转波片偏振检测装置进行参数校准

阅读:1

Abstract

When measuring Stokes parameters using the rotating wave plate method, the angle error of the polarizer's light transmission axis, the azimuth error of the wave plate's fast axis, and the phase delay error are key factors restricting accuracy. To address the existing calibration methods' insufficient accuracy and incomplete consideration of the error parameters, this study constructed an error-transfer analytical model for an in-depth analysis of the principle of measuring Stokes parameters using the rotating wave plate method. It also clarified the quantitative parameter relationship between the measurement, wave plate, and polarizer errors. A device parameter calibration scheme using multi-angle polarized light (horizontally linearly polarized, [1,1,0,0](T), and 45° linearly polarized, [1,0,1,0](T)) was further proposed, and by using the deviation between the theoretical response of the standard incident light and the actual measurement data, an error equation was established to solve the device parameter error and precisely calibrate the polarization detection device. The experimental results show that after using this method, the calibration error of the Stokes parameters decreased from 4.83% to within 0.46%, significantly overcoming the traditional methods' limitations regarding incomplete consideration of the error parameters and accuracy improvement, providing a more concise and reliable method for high-precision polarization measurement.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。