Precision of a new ocular biometer in children and comparison with IOLMaster

新型眼部生物测量仪在儿童中的精度及其与IOLMaster的比较

阅读:5

Abstract

To assess the repeatability and reproducibility of AL-Scan in agreement with those by the IOLMaster in healthy children, two skilled operators measured ocular parameters in 58 children. The parameters included keratometry (K) values, anterior chamber depth (ACD), axial length (AL), central corneal thickness (CCT), pupil diameter (PD), and corneal diameter (CD). The cohort comprised of 32 boys and 26 girls. The AL-Scan measurements showed high repeatability, as the test-retest repeatability (TRT) values of AL, CCT, ACD, Kf, Ks, Km, CD, and PD were 0.09 mm, 5.1 μm, 0.04 mm, 0.28 D, 0.24 D, 0.21 D, 0.39 mm, and 0.22 mm, respectively. The within-subject coefficient of variation (CoV) was low (<0.35%) and the intraclass correlation coefficients (ICC) of all parameters were >0.85. The interobserver reproducibility was excellent with low values of TRT and ICC > 0.95. The CoV of AL, CCT, ACD, and K was <0.22%. The 95% limits of agreement between the AL-Scan and the IOLMaster were narrow for all parameters except for CD. The repeatability and reproducibility of the new biometer, Al-Scan, was excellent for all parameters and can be routinely used in children to measure the biometric values.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。