The Coefficient of Variation of Step Time Can Overestimate Gait Abnormality: Test-Retest Reliability of Gait-Related Parameters Obtained with a Tri-Axial Accelerometer in Healthy Subjects

步时变异系数可能高估步态异常:健康受试者使用三轴加速度计获得的步态相关参数的重测信度

阅读:1

Abstract

The aim of this study was to investigate whether variation in gait-related parameters among healthy participants could help detect gait abnormalities. In total, 36 participants (21 men, 15 women; mean age, 35.7 ± 9.9 years) performed a 10-m walk six times while wearing a tri-axial accelerometer fixed at the L3 level. A second walk was performed ≥1 month after the first (mean interval, 49.6 ± 7.6 days). From each 10-m data set, the following nine gait-related parameters were automatically calculated: assessment time, number of steps, stride time, cadence, ground force reaction, step time, coefficient of variation (CV) of step time, velocity, and step length. Six repeated measurement values were averaged for each gait parameter. In addition, for each gait parameter, the difference between the first and second assessments was statistically examined, and the intraclass correlation coefficient (ICC) was calculated with the level of significance set at p < 0.05. Only the CV of step time showed a significant difference between the first and second assessments (p = 0.0188). The CV of step time also showed the lowest ICC, at <0.50 (0.425), among all parameters. Test-retest results of gait assessment using a tri-axial accelerometer showed sufficient reproducibility in terms of the clinical evaluation of all parameters except the CV of step time.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。