Optimizing image guidance frequency and implications on margins for gynecologic malignancies

优化图像引导频率及其对妇科恶性肿瘤手术切缘的影响

阅读:1

Abstract

BACKGROUND: To analyze setup deviations using daily megavoltage computed tomography (MVCT) and to evaluate three MVCT frequency reducing protocols for gynecologic cancer patients treated with helical tomotherapy. METHODS: We recorded the setup errors of 56 patients with gynecological cancer observed throughout their whole course by matching their daily MVCT with the planning CT. Systematic and random errors were calculated on a patient and population basis. We defined three different protocols corresponding to MVCTs from the first five fractions (FFF), the first ten fractions (FTF) or from the first and third weeks (505). We compared theoretical. setup errors calculated using these 5 or 10 early MVCT scans with the actual errors found with the remaining fractions to to analyze the residual deviations. RESULTS: The total systematic (random) deviations had means of -2.0 (3.8)mm, 0.5 (3.4)mm, 0.5 (6.1)mm and -0.5° (0.9°) in vertical (V), longitudinal (LO), lateral (LA), and roll (R) directions, respectively. The proposed three MVCT protocols resulted in minor residual deviations. In all three protocols, 95% of all calculated residual deviations were less than or equal to 5 mm in all 3 directions. When examining the additional minimal CTV-PTV setup margins that were calculated based on these residual deviations, the 505 protocol would have allowed smaller margins than the FFF and FTF protocol, particularly in the V direction. CONCLUSIONS: For patients with gynecologic cancer, the 505-protocol led to the lowest residual deviations and therefore might offer the best approach in reducing the frequency of pre-treatment MVCTs.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。