Safe Hb Concentration Measurement during Bladder Irrigation Using Artificial Intelligence

利用人工智能在膀胱灌洗过程中安全测量血红蛋白浓度

阅读:1

Abstract

We have developed a sensor for monitoring the hemoglobin (Hb) concentration in the effluent of a continuous bladder irrigation. The Hb concentration measurement is based on light absorption within a fixed measuring distance. The light frequency used is selected so that both arterial and venous Hb are equally detected. The sensor allows the measurement of the Hb concentration up to a maximum value of 3.2 g/dL (equivalent to ≈20% blood concentration). Since bubble formation in the outflow tract cannot be avoided with current irrigation systems, a neural network is implemented that can robustly detect air bubbles within the measurement section. The network considers both optical and temporal features and is able to effectively safeguard the measurement process. The sensor supports the use of different irrigants (salt and electrolyte-free solutions) as well as measurement through glass shielding. The sensor can be used in a non-invasive way with current irrigation systems. The sensor is positively tested in a clinical study.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。