Quality assurance of non-coplanar, volumetric-modulated arc therapy employing a C-arm linear accelerator, featuring continuous patient couch rotation

采用C臂直线加速器进行非共面容积调强弧形治疗的质量保证,该治疗具有连续的患者治疗床旋转功能。

阅读:2

Abstract

PURPOSE: To perform quality assurance of non-coplanar, volumetric-modulated arc therapy featuring continuous couch rotation (CCR-VMAT) using a C-arm linear accelerator. METHODS: We planned and delivered CCR-VMAT using the TrueBeam Developer Mode. Treatment plans were created for both a C-shaped phantom and five prostate cancer patients using seven CCR trajectories that lacked collisions; we used RayStation software (ver. 4.7) to this end. Subsequently, verification plans were generated. The mean absolute error (MAE) between the center of an MV-imaged steel ball and the radiation field was calculated using the Winston-Lutz test. The MAEs between planned and actual irradiation values were also calculated from trajectory logs. In addition, correlation coefficients (r values) among the MAEs of gantry angle, couch angle, and multi-leaf collimator (MLC) position, and mechanical parameters including gantry speed, couch speed, MLC speed, and beam output, were estimated. The dosimetric accuracies of planned and measured values were also assessed using ArcCHECK. RESULTS: The MAEs ±2 standard deviations as revealed by the Winston-Lutz test for all trajectories were 0.3 ± 0.3 mm in two dimensions. The MAEs of the gantry, couch, and MLC positions calculated from all trajectory logs were within 0.04°, 0.08°, and 0.02 mm, respectively. Deviations in the couch angle (r = 0.98, p < 0.05) and MLC position (r = 0.86, p < 0.05) increased significantly with speed. The MAE of the beam output error was less than 0.01 MU. The mean gamma passing rate ± 2 SD (range) of the 3%/3 mm, 3%/1 mm, and 5%/1 mm was 98.1 ± 1.9% (95.7-99.6%), 87.2 ± 2.8% (80.2-96.7%), and 96.3 ± 2.8% (93.9-99.6%), respectively. CONCLUSIONS: CCR-VMAT delivered via the TrueBeam Developer Mode was associated with high-level geometric and mechanical accuracy, thus affording to high dosimetric accuracy. The CCR-VMAT performance was stable regardless of the trajectory chosen.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。