Clinical application of multi-material artifact reduction (MMAR) technique in Revolution CT to reduce metallic dental artifacts

Revolution CT中多材料伪影减少(MMAR)技术在减少金属牙科伪影方面的临床应用

阅读:2

Abstract

BACKGROUND: This study aimed to explore the performance of Revolution CT virtual monoenergetic images (VMI) combined with the multi-material artifact reduction (MMAR) technique in reducing metal artifacts in oral and maxillofacial imaging. RESULTS: There were significant differences in image quality scores between VMI + MMAR images and VMI+MARS (multiple artifact reduction system) images at each monochromatic energy level (p = 0.000). Compared with the MARS technology, the MMAR technology further reduced metal artifacts and improved the image quality. At VMI(90 keV) and VMI(110 keV), the SD, CNR, and AI in the Revolution CT group were significantly lower than in the Discovery CT, but no significant differences in these parameters were found between two groups at VMI(50 keV), VMI(70 keV), and VMI(130 keV) (p > 0.05). The attenuation was comparable between two groups at any energy level (p > 0.05). CONCLUSIONS: Compared with the MARS reconstruction technique of Discovery CT, the MMAR technique of Revolution CT is better to reduce the artifacts of dental implants in oral and maxillofacial imaging, which improves the image quality and the diagnostic value of surrounding soft tissues.