Analytical performance evaluation of five RT-PCR kits for severe acute respiratory syndrome coronavirus 2

对五种严重急性呼吸综合征冠状病毒2的RT-PCR试剂盒进行分析性能评估

阅读:2

Abstract

BACKGROUND: We aimed to evaluate the analytical performance of five commercial RT-PCR kits (Genekey, Daan, BioGerm, Liferiver, and Yaneng) commonly used in China, since such comparison data are lacking. METHODS: A total of 20 COVID-19 confirmed patients and 30 negative nasopharyngeal swab specimens were analyzed by five kits. The detection ability of five RT-PCR kits was evaluated with 5 concentration gradients diluted by a single positive sample. The limit of detection was evaluated by N gene fragment solid standard. Two positive clinical specimens were used to evaluate the repeatability and imprecision. Finally, we used six human coronaviruses plasmid and four respiratory pathogens plasmid to check for cross-reactivity. RESULTS: The positive detection rate was 100% for Genekey, Daan, and BioGerm,and 90% for Liferiver and Yaneng in 20 clinical SARS-CoV-2 infection. The coincidence rate of five kits in 10 negative samples was 100%. The detection rate of target genes for Daan, BioGerm, Liferiver, and Yaneng was 100% from Level 1 to Level 3. In Level 4, only Daan detection rate was 100%. In Level 5, five kits presented poor positive rate. The limit of detection declared by each manufacturer was verified. The repeatability for target genes was less than 5% and so did the total imprecision. There is no cross-reactivity of five kits with six human coronaviruses and four respiratory pathogens for ORF1ab and N gene. CONCLUSIONS: Five RT-PCR kits assessed in this study showed acceptable analytical performance characteristics and are useful tools for the routine diagnosis of SARS-CoV-2.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。