Solid-Contact Ion-Selective Electrodes for Histamine Determination

用于测定组胺的固态接触离子选择性电极

阅读:1

Abstract

Solid-contact ion-selective electrodes for histamine (HA) determination were fabricated and studied. Gold wire (0.5 mm diameter) was coated with poly(3,4-ethlenedioxythiophene) doped with poly(styrenesulfonate) (PEDOT:PSS) as a solid conductive layer. The polyvinyl chloride matrix embedded with 5,10,15,20-tetraphenyl(porphyrinato)iron(iii) chloride as an ionophore, 2-nitrophenyloctyl ether as a plasticizer and potassium tetrakis(p-chlorophenyl) borate as an ion exchanger was used to cover the PEDOT:PSS layer as a selective membrane. The characteristics of the HA electrodes were also investigated. The detection limit of 8.58 × 10(-6) M, the fast response time of less than 5 s, the good reproducibility, the long-term stability and the selectivity in the presence of common interferences in biological fluids were satisfactory. The electrode also performed stably in the pH range of 7-8 and the temperature range of 35-41 °C. Additionally, the recovery rate of 99.7% in artificial cerebrospinal fluid showed the potential for the electrode to be used in biological applications.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。