Image Quality Analysis of Photon-Counting CT Compared with Dual-Source CT: A Phantom Study for Chest CT Examinations

光子计数CT与双源CT图像质量分析:胸部CT检查的体模研究

阅读:2

Abstract

A comparison was made between the image quality of a photon-counting CT (PCCT) and a dual-source CT (DSCT). The evaluation of image quality was performed using a Catphan CT phantom, and the physical metrics, such as the noise power spectrum and task transfer function, were measured for both PCCT and DSCT at three CT dose indices (1, 5 and 10 mGy). Polyenergetic and virtual monoenergetic reconstructions were used to evaluate the performance differences by simulating a Gaussian spot with a radius of 5 mm and calculating the detectability index. The highest iterative reconstruction level was able to decrease the noise by about 70% compared with the filtered back projection using a parenchyma reconstruction kernel. The PCCT task transfer functions remained constant, while those of the DSCT increased with the reconstruction strength level. At monoenergetic 70 keV, a 50% decrease in noise was observed for DSCT with image smoothing, while PCCT had the same 50% decrease in noise without any smoothing. The PCCT detectability index at a reconstruction strength level of two was equivalent to the highest level of ADMIRE 5 for DSCT. The PCCT showed its superiority over the DSCT, especially for lung nodule detection.