Spectral imaging using clinical megavoltage beams and a novel multi-layer imager

利用临床兆伏级射线和新型多层成像器进行光谱成像

阅读:2

Abstract

We assess the feasibility of clinical megavoltage (MV) spectral imaging for material and bone separation with a novel multi-layer imager (MLI) prototype. The MLI provides higher detective quantum efficiency and lower noise than conventional electronic portal imagers. Simulated experiments were performed using a validated Monte Carlo model of the MLI to estimate energy absorption and energy separation between the MLI components. Material separation was evaluated experimentally using solid water and aluminum (Al), copper (Cu) and gold (Au) for 2.5 MV, 6 MV and 6 MV flattening filter free (FFF) clinical photon beams. An anthropomorphic phantom with implanted gold fiducials was utilized to further demonstrate bone/gold separation. Weighted subtraction imaging was employed for material and bone separation. The weighting factor (w) was iteratively estimated, with the optimal w value determined by minimization of the relative signal difference ([Formula: see text]) and signal-difference-to-noise ratio (SDNR) between material (or bone) and the background. Energy separation between layers of the MLI was mainly the result of beam hardening between components with an average energy separation between 34 and 47 keV depending on the x-ray beam energy. The minimum average energy of the detected spectrum in the phosphor layer was 123 keV in the top layer of the MLI with the 2.5 MV beam. The w values that minimized [Formula: see text] and SDNR for Al, Cu and Au were 0.89, 0.76 and 0.64 for 2.5 MV; for 6 MV FFF, w was 0.98, 0.93 and 0.77 respectively. Bone suppression in the anthropomorphic phantom resulted in improved visibility of the gold fiducials with the 2.5 MV beam. Optimization of the MLI design is required to achieve optimal separation at clinical MV beam energies.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。