Fracture strength analysis of titanium insert-reinforced zirconia abutments according to the axial height of the titanium insert with an internal connection

根据钛嵌体轴向高度对带内连接的钛嵌体增强氧化锆基台的断裂强度进行分析

阅读:2

Abstract

This study aimed to analyze fracture strength in vitro by varying the axial height of the titanium insert and the labial height of the zirconia abutment in an internal connection implant to identify the titanium insert axial height with optimal mechanical stability. Sixty implants with an internal connection system were used. Two-piece zirconia abutments were used with the titanium inserts. Combinations of different titanium insert axial heights (mm) and zirconia abutment labial heights (mm) constituted five groups: Gr1 (1-3), Gr2 (3-3), Gr3 (3-5), Gr4 (5-3), and Gr5 (5-5). After thermocycling, a fracture load test was performed with a universal testing machine. The initial deformation load and the fracture load were measured and analyzed. The fractured surface and cross-section of the specimens were examined by scanning electron microscopy (SEM). The groups of titanium inserts with axial heights of 3 mm and 5 mm showed significantly greater initial deformation load and fracture load than the group with an axial height of 1 mm (p < 0.05), but there was no significant difference between the two groups with axial heights of 3 mm and 5 mm. The labial height of the zirconia abutment had no significant influence on the initial deformation load and fracture load. In some specimens in Gr4 and Gr5, cracking or bending of the titanium insert and abutment screw was observed on SEM. The axial height of the titanium insert should be designed to not be less than 3 mm to increase the fracture strength and promote the long-term stability of implants.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。