Point-of-Care Cerebrospinal Fluid Detection

即时脑脊液检测

阅读:9
作者:Ashley E Kita, Daniel W Bradbury, Zachary D Taylor, Daniel T Kamei, Maie A St John

Conclusion

Herein we detail the creation and initial validation of the first point-of-care cerebrospinal fluid detection device. This device is a feasible method to more efficiently and cost-effectively identify cerebrospinal fluid leaks, minimize costs, and improve patient outcomes.

Methods

Tests with known concentrations of resuspended beta-trace protein and the contents of discarded lumbar drains (presumed to contain cerebrospinal fluid) were performed to validate our novel device.

Objective

A cerebrospinal fluid leak is one of the most serious complications in otolaryngology. It may occur as a result of injury to the skull base, typically traumatic or iatrogenic. While the presence of a leak is often discerned in the emergent setting, distinguishing normal secretions from those containing cerebrospinal fluid can be difficult during postoperative visits in the clinic. As most current laboratory-based assays are labor intensive and require several days to result, we aim to develop a more user-friendly and rapid point-of-care cerebrospinal fluid detection device. Study design: Our laboratory developed a barcode-style lateral-flow immunoassay utilizing antibodies for beta-trace protein, a protein abundant in and specific for cerebrospinal fluid, with a concentration of 1.3 mg/L delineating a positive result. Setting: Tertiary medical center. Subjects and

Results

Our results demonstrate the ability of our device to semiquantitatively identify concentrations of beta-trace protein from 0.3-90 mg/L, which is within the required range to diagnose a leak, thus making beta-trace protein an excellent target for rapid clinical detection.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。