Technical considerations for medical device manufacturers when designing gastrostomy tubes (G-tubes) using the new ISO 80369-3 connector

医疗器械制造商在使用新型 ISO 80369-3 连接器设计胃造瘘管(G 管)时需要考虑的技术因素

阅读:2

Abstract

BACKGROUND: Gastrostomy tubes (G-tubes) are typically used when people cannot eat food by mouth. The connector section that allows G-tubes to connect to other devices, such as feeding sets or syringes, has been modified on some of the devices to reduce misconnections in hospital settings. The narrow internal diameter of the new connector, standardized under ISO 80369-3, has caused some users to express concern about a reduced flow rate. Previous studies performed on commercial devices determined that it was not conclusive how much the ISO 80369-3 connector contributed towards the reduced flow rate, because when manufacturers designed these new connector-based devices, they often changed other geometric variables (such as distal tube diameter, or length) at the same time. Thus, it became difficult isolating the effect of the connector from other geometric variables. METHOD: The key objective of this study was to investigate how different design variables impacted the flow rate through the G-tubes. 3D-printed devices were used to assess the geometric parameters in a systematic manner. Commercial diets and Newtonian analog fluids with matched viscosities were used for testing. RESULTS: The flow path length of the "transition section" encompassing the standardized ISO 80369-3 connector in the new devices was found to cause reduced flow. Additionally, results showed that a shortened (≤ 10 mm) transition section, along with a 10% increase in the distal inner diameter of large bore devices (e.g., 24 Fr), can restore flow rates to levels consistent with the previous devices prior to the connector standardization. CONCLUSIONS: The strategy for restoring flow rates to previous levels may help alleviate concerns raised by multiple stakeholders such as health care professionals, patients, caregivers and device manufacturers. In addition, the approach proposed here can be used as a tool for designing future G-tube devices.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。