Label-Free Electrochemical Immunosensor Based on β-Cyclodextrin-Functionalized Helical Carbon Nanotube and Ionic Liquid Containing Ferrocene and Aldehyde Groups

基于β-环糊精功能化螺旋碳纳米管和含二茂铁和醛基离子液体的无标记电化学免疫传感器

阅读:2

Abstract

We fabricated an electrochemical immunosensor for the detection of cardiac troponin I using β-cyclodextrin-functionalized helical carbon nanotube and ionic liquid functionalized with ferrocene and aldehyde groups. β-Cyclodextrin-functionalized helical carbon nanotube was first modified on the electrode surface. Then, ferrocene- and aldehyde-functionalized ionic liquid was modified on the surface of the electrode through host-guest interaction, resulting in an interface with aldehyde groups. The aldehyde groups attached to the ionic liquid capture antibody directly, which simplifies the fabrication of immunosensor. Because of the use of ionic liquid and helical carbon nanotube, the conductivity of the sensing interface was improved. Thus, the sensitivity of the fabricated immunosensor was increased. The immunosensor for cardiac troponin I shows a linear range from 0.05 to 20 ng mL(-1) with a detection limit of 0.04 ng mL(-1) (S/N = 3).