Temperature loss by ventilation in a calorimetric bench model

量热台模型中通风造成的温度损失

阅读:2

Abstract

In intensive care medicine heat moisture exchangers are standard tools to warm and humidify ventilation gases in order to prevent temperature loss of patients or airway epithelia damage. Despite being at risk of hypothermia especially after trauma, intubated emergency medicine patients are often ventilated with dry and in winter probably cold ventilation gases. We tried to assess the amount of temperature-loss due to ventilation with cold, dry medical oxygen in comparison to ventilation with warm and humidified oxygen. We ventilated a 50-kg water-dummy representing the calorimetric capacity of a 60-kg patient over a period of 2 hours (tidal volume 6.6 mL/kg = 400 mL; respiratory rate 13/min). Our formal null-hypothesis was that there would be no differences in temperature loss in a 50 kg water-dummy between ventilation with dry oxygen at 10°C vs. ventilation with humidified oxygen at 43°C. After 2 hours the temperature in the water-dummy using cold and dry oxygen was 29.7 ± 0.1°C compared to 30.4 ± 0.1°C using warm and humidified oxygen. This difference in cooling rates between both ventilation attempts of 0.7 ± 0.1°C after 2 hours represents an increased cooling rate of ~0.35°C per hour. Ventilation with cool, dry oxygen using an automated transport ventilator resulted in a 0.35°C faster cooling rate per hour than ventilation with warm humidified oxygen in a bench model simulating calorimetric features of a 60-kg human body.