Polarization modulated spectroscopic ellipsometry-based surface plasmon resonance biosensor for E. coli K12 detection

基于偏振调制光谱椭偏仪的表面等离子体共振生物传感器用于大肠杆菌K12的检测

阅读:2

Abstract

In this work, we report on the application of the polarization modulated spectroscopic ellipsometry-based surface plasmon resonance method for sensitive detection of microorganisms in Kretschmann configuration. So far, rotating analyzer and single wavelength polarization modulation methods have widely been investigated for phase sensitive surface plasmon resonance measurement. In this study, a much simpler optical setup relying on fast electro-optic phase modulator crystals is introduced for bacteria detection. A beta barium borate crystal connected to a function generator is adapted for generating phase shifts in the millisecond regime to extract the ellipsometric angles ( Ψ and Δ ) under the surface plasmon resonance condition. For detection, the gold surface was functionalized with anti-Escherichia coli antibodies, and E. coli K12 was attached to them. We show that polarization modulated spectroscopic ellipsometry achieves a refractive index resolution in the order of 10-5 RIU, and a limit of detection of 102 CFU/mL for E. coli K12 which is compatible with other surface plasmon resonance based phase sensitive methods with more complex detection concepts. As a follow-up step, an optical model can be developed to enhance this biosensor's performance, and applications for sorting and detecting other biological targets will be investigated.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。