Regenerative Strategy of Gold Electrodes for Long-Term Reuse of Electrochemical Biosensors

用于电化学生物传感器长期重复使用的金电极再生策略

阅读:2

Abstract

Gold is of considerable interest for electrochemical active surfaces because thiol-modified chemicals and biomolecules can be easily immobilized with a simple procedure. However, most gold surfaces are damaged with repetitive measurements, so they are difficult to reuse. Here we demonstrate a novel electrochemical cleaning method of gold surfaces to reuse electrodes with a simple protocol that is easy and nontoxic. This electrochemical cleaning consists of two steps by using different solutions. The 1st step is a cyclic voltammetry sweep using a very low concentration of sulfuric acid, and the 2nd step is a cyclic voltammetry sweep using potassium ferricyanide. Different cleaning methods were also considered for comparison. Consequently, after assembling and desorption of the cell and antigen, the changes in gold electrode performance, as immunosensor and cytosensor, were investigated by electrochemical impedance and cyclic voltammetry. It was found that repetitive measurement is possible until five times while maintaining the reproducibility. It is believed that this method is capable of enabling reuse of gold electrodes and can be used for long-term and accurate monitoring of biological effects, especially at a low cost.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。