Analytical Performance of the Factory-Calibrated Flash Glucose Monitoring System FreeStyle Libre2TM in Healthy Women

工厂校准的闪光血糖监测系统 FreeStyle Libre2TM 在健康女性中的分析性能

阅读:5
作者:Zhuoxiu Jin, Alice E Thackray, James A King, Kevin Deighton, Melanie J Davies, David J Stensel

Abstract

Continuous glucose monitoring (CGM) is used clinically and for research purposes to capture glycaemic profiles. The accuracy of CGM among healthy populations has not been widely assessed. This study assessed agreement between glucose concentrations obtained from venous plasma and from CGM (FreeStyle Libre2TM, Abbott Diabetes Care, Witney, UK) in healthy women. Glucose concentrations were assessed after fasting and every 15 min after a standardized breakfast over a 4-h lab period. Accuracy of CGM was determined by Bland-Altman plot, 15/15% sensor agreement analysis, Clarke error grid analysis (EGA) and mean absolute relative difference (MARD). In all, 429 valid CGM readings with paired venous plasma glucose (VPG) values were obtained from 29 healthy women. Mean CGM readings were 1.14 mmol/L (95% CI: 0.97 to 1.30 mmol/L, p < 0.001) higher than VPG concentrations. Ratio 95% limits of agreement were from 0.68 to 2.20, and a proportional bias (slope: 0.22) was reported. Additionally, 45% of the CGM readings were within ±0.83 mmol/L (±15 mg/dL) or ±15% of VPG, while 85.3% were within EGA Zones A + B (clinically acceptable). MARD was 27.5% (95% CI: 20.8, 34.2%), with higher MARD values in the hypoglycaemia range and when VPG concentrations were falling. The FreeStyle Libre2TM CGM system tends to overestimate glucose concentrations compared to venous plasma samples in healthy women, especially during hypoglycaemia and during glycaemic swings.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。