AgNPs@CeO(2)/Nafion Nanocomposite-Modified Electrode for the Sensitive Detection of Trace Lead (II) in Water Samples

用于灵敏检测水样中痕量铅(II)的AgNPs@CeO(2)/Nafion纳米复合修饰电极

阅读:1

Abstract

Excessive levels of heavy metal pollutants in the environment pose significant threats to human health and ecosystem stability. Consequently, the accurate and rapid detection of heavy metal ions is critically important. A AgNPs@CeO(2)/Nafion composite was prepared by dispersing nano-ceria (CeO(2)) in a Nafion solution and incorporating silver nanoparticles (AgNPs). The morphology, microstructure, and electrochemical properties of the modified electrode materials were systematically characterized using scanning electron microscopy (SEM), energy-dispersive X-ray spectroscopy (EDX), X-ray diffraction (XRD), and cyclic voltammetry (CV). By leveraging the oxygen vacancies and high electron transfer efficiency of CeO(2), the strong adsorption capacity of Nafion, and the superior conductivity of AgNPs, an AgNPs@CeO(2)/Nafion/GCE electrochemical sensor was developed. Under optimized conditions, trace Pb(2+) in water was detected using square wave anodic stripping voltammetry (SWASV). The sensor demonstrated a linear response for Pb(2+) within the concentration range of 1-100 μg·L(-1), with a detection limit of 0.17 μg·L(-1) (S/N = 3). When applied to real water samples, the method achieved recovery rates between 93.7% and 110.3%, validating its reliability and practical applicability.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。