Preliminary results of implantation in animal model and osteoblast culture evaluation of prototypes of biomimetic multispiked connecting scaffold for noncemented stemless resurfacing hip arthroplasty endoprostheses

非骨水泥型无柄髋关节置换假体仿生多钉连接支架原型动物模型植入及成骨细胞培养评价的初步结果

阅读:7
作者:Ryszard Uklejewski, Piotr Rogala, Mariusz Winiecki, Andrzej Kędzia, Piotr Ruszkowski

Abstract

We present the new fixation method for RHA (resurfacing hip arthroplasty) endoprostheses by means of the biomimetic multispiked connecting scaffold (MSC-Scaffold). Such connecting scaffold can generate new type of RHA endoprostheses, that is stemless and fixed entirely without cement. The preprototypes of this MSC-Scaffold were manufactured with modern additive laser additive technology (SLM). The pilot surgical implantations in animal model (two laboratory swine) of MSC-Scaffold preprototypes have showed after two months neither implant loosening, migration, and nor other early complications. From the results of performed histopathological evaluation of the periscaffold spikes bone tissue and 10-day culture of human osteoblasts (NHOst) we can conclude that (1) the scaffolding effect was obtained and (2) to improve the osseointegration of the scaffold spikes, their material surface should be physicochemically modified (e.g., with hydroxyapatite). Some histopathological findings in the periscaffold domain near the MSC-Scaffold spikes bases (fibrous connective tissue and metallic particles near the MSC-Scaffold spikes bases edges) prompt considering the necessity to optimize the design of the MSC-Scaffold in the regions of its interspike space near the spikes bases edges, to provide more room for new bone formation in this region and for indispensable post-processing (glass pearl blasting) after the SLM manufacturing.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。