Nitric Oxide Analysis Down to ppt Levels by Optical-Feedback Cavity-Enhanced Absorption Spectroscopy

利用光反馈腔增强吸收光谱法分析低至ppt级的一氧化氮

阅读:1

Abstract

Monitoring nitric oxide at the trace level is required in a large range of applications. We report on a trace gas analyzer optimized for nitric oxide measurements by Optical Feedback Cavity Enhanced Absorption Spectroscopy with an interband cascade laser at 5.3 µm. The short response time of the instrument allows for reaching the level of 50 ppt in only 180 ms. Its stability enables averaging up to 12 min to reach a detection limit of 0.9 ppt. Absolute concentration calibration requires to account for the optical saturation effect that results from the intense absorption line intensity addressed here, in the mid infrared region, in contrast to instruments that are operating in the near infrared region.